From your knowledge of expected concentrations, use the table below to determine which
settings and solutions are appropriate:
Use Total LR or Strong LR for waters alkalinity in the 10 - 500 mg/L range.
Use Total HR or Strong HR for waters alkalinity in the 400 - 4000 mg/L range.
Use the appropriate titrant for each range.
Also, the pump must be recalibrated when the measurement range is changed (see table above).
Alkalinity type
SETUP CONFIGURATION MENU
Options: mg/L, meq/L.
Press the corresponding function key to change the
option.
Concentration unit
The titrator’s setup configuration menu may be accessed from the pH or titration screens by
pressing the MENU key, then Setup.
A list of setup parameters will be displayed with currently configured setting.
While in the setup menu it is possible to modify the instrument’s operation parameters. The
ARROW keys permit the user to scroll the setup parameters.
Press HELP to view the contextual help.
Press ESC to return to the main screen.
Options: Total LR (total alkalinity low range)
Total HR (total alkalinity high range)
Strong LR (strong alkalinity low range)
Strong HR (strong alkalinity high range)
Press the corresponding function key to change the
option.