Hanna-instruments HI 96104 Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 28
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
18
CYANURIC ACID MEASUREMENT
Fill the cuvette with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
Fill a beaker up to the 25 mL mark with
the sample, add the content of one packet
of HI 93722-0 reagent and swirl gently
to mix. This is the reacted sample.
Fill a second cuvette with 10 mL of the
reacted sample, up to the mark, and
replace the cap.
Place the second cuvette into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
Vista de pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 27 28

Comentarios a estos manuales

Sin comentarios