Remove the cuvette.
Add the content of one packet of
HI 93708-0 reagent. Replace the cap
and shake gently.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 10 minutes then
just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
nitrite on the LCD.
To convert the NO
2
-
ion concentration to
NO
2
-
-N concentration, divide the reading
by a factor of 3.29.
To convert the NO
2
-
ion to sodium nitrite
NaNO
2
concentration, multiply the reading
by a factor of 1.5.
INTERFERENCE
HI 96708 will not recognize any glycol based samples.