5
SPECIFICATIONS
PRECISION AND ACCURACY
Range 0 to 150 mg/L
Resolution 1 mg/L
Accuracy ±4 mg/L ±4% of reading @ 25°C
Typical EMC Deviation ±1 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell with narrow band interference filter @ 575nm
Method Adaptation of the Ferrous Sulfate method. The reaction between nitrite
and the reagent causes a greenish-brown tint in the sample.
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in
measurement mode
;
after 1 hour of non-use in
calibration mode
;
with last reading reminder.
Dimensions 192 x 102 x 67 mm (7.6 x 4 x 2.6")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity/test
HI 93708-0 Powder Reagent 1 packet
PrecisionPrecision
PrecisionPrecision
Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
AccuracyAccuracy
AccuracyAccuracy
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of
75 mg/L chromium VI and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 4 mg/L.