15
VALIDATION PROCEDURE
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
• Turn the meter on by pressing ON/OFF.
• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
• Place the CAL CHECK
TM
Standard
HI 96709-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
INTERFERENCES
• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” asks for instrument
zeroing.