CALIBRATION PROCEDURE
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or
ON/OFF keys.
Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Turn the meter on by pressing ON/OFF.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” asks for instrument
zeroing.
Press and hold CAL CHECK for three
seconds to enter
calibration mode
. The
display will show “CAL” during
calibration procedure. The blinking
“ZERO” asks for instrument zeroing.
Place the CAL CHECK
TM
Standard
HI 96709-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.