TOTAL CHLORINE MEASUREMENT
Fill the cuvette up to the mark with 10
mL of unreacted sample and replace the
cap.
Place the cuvette into the holder and
ensure that the notch on the cap is posi-
tioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
Remove the cuvette and add one packet
of HI 93711-0 reagent. Replace the cap
and shake gently for 20 seconds (or
2 minutes in case of seawater analysis).
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.