3.2 PROBE SPECIFICATIONS
Non-logging Probe Logging Probe
Sample Environment Fresh, brackish, seawater
Waterproof protection IP68
Computer Interface NA USB PC (HI 76982910)
Internal Battery Type NA 4 X 1.5V Size AA Alkaline
Typical Battery Life NA See below
Memory NA 140,000 measurements
(single parameter logged)
35,000 measurements
(all parameters logged)
Operating Temperature -5 to 55° C *
Storage Temperature -20 to 70° C
Maximum Depth 20 m (66 ft.) *
Dimensions
HI 7609829HI 7609829
HI 7609829HI 7609829
HI 7609829 342mm (13.5”),
HI 7629829HI 7629829
HI 7629829HI 7629829
HI 7629829 442mm (17.4"),
(without cable) dia=46 mm (1.8”) dia=46 mm (1.8")
HI 7619829HI 7619829
HI 7619829HI 7619829
HI 7619829 382 mm (15.1”),
HI 7639829HI 7639829
HI 7639829HI 7639829
HI 7639829 482 mm (19.0"),
dia=46 mm (1.8”) dia=46 mm (1.8")
Weight
HI 7609829HI 7609829
HI 7609829HI 7609829
HI 7609829 570g (20.1 oz.)
HI 7629829HI 7629829
HI 7629829HI 7629829
HI 7629829 775g (27.3 oz.)
(with batteries and sensors)
HI 7619829HI 7619829
HI 7619829HI 7619829
HI 7619829 650g (22.9 oz.)
HI 7639829HI 7639829
HI 7639829HI 7639829
HI 7639829 819g (28.9oz.)
Cable Specification Multistrand-multiconductor shielded cable with internal strength
member rated for 68 kg (150 lb) intermittent use
Wetted Materials Body: ABS
Threads: Nylon
Shield: ABS/ 316 SS
Temp probe: 316 SS
O-rings: EPDM
* Reduced for ISE sensors
LOGGING PROBE BATTERY LIFE*
Interval All channels logging All channels logging
(no averaging) (10 sample averaging)
1 - 5 sec 72 hours 72 hours
1 min 22 days 11 days
10 min 70 days 65 days
* Continuous logging is contingent on availability of log memory