7.1 QUICK CALIBRATION
The quick calibration method provides a quick single point calibration for pH,
conductivity and dissolved oxygen sensors. HI 9828-25 calibration solution is
used for both pH and conductivity.
Fill the calibration beaker 2/3 full with HI 9828-25 calibration solution.
Slowly place the sensors into the solution and
dislodge bubbles that may adhere to the sensors.
Screw the calibration beaker completely on the
probe body. Some solution may overflow.
Wait a few minutes for the system to stabilize.
From the “Calibration” menu select “Quick
calibration”.
A three item calibration menu will appear (pH,
Conductivity and Dissolved oxygen) and “pH”
will start to blink along with the “Not ready
message.
When the pH signal is stable, the “Ready
message appears. Press <Confirm> to store the
calibration data.
The “Storing” message will appear as the
calibration proceeds to the next sensor. A
checkmark will appear in the box next to “pH” to
indicate a successful calibration.
NoteNote
NoteNote
Note To bypass any of the calibrations press
<Skip> to move to the next sensor in the
quick calibration menu.
If the pH sensor is not installed the message
“pH sensor not installed! Skip to
conductivity calibration” will appear.
Following the pH calibration, “Conductivity” will start to blink along with the
“Not ready” message.
When the measurement is stable, “Ready” appears. Press <Confirm> to store
the calibration data and the “Storing” message will appear.
NoteNote
NoteNote
Note If EC calibration is not required, skip to
the D.O. quick calibration by pressing the
<Skip> softkey.
The message “Empty the beaker.” will appear.