Hanna-instruments HI 9829 Manual de usuario Pagina 37

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 36
37
Unscrew the calibration beaker and empty the solution.
Shake any remaining liquid off the probe and beaker. No droplets
should remain on the D.O. sensor membrane.
NoteNote
NoteNote
Note Do not attempt to dry wipe the D.O. sensor as damage to the mem
brane may occur.
Screw the empty calibration beaker on the probe body. The beaker should not
be dry.
Press <Accept> to close the displayed message.
When the measurement is stable, “Ready
appears. Press <Confirm> to store the
calibration data and the “Storing” message will
appear.
Press <OK> to return to “Calibration” menu.
NoteNote
NoteNote
Note To quit the quick calibration procedure, press ESC at any time.
After every calibration the quick calibration window will show a check mark in
the box next to the calibrated parameter.
Vista de pagina 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 103 104

Comentarios a estos manuales

Sin comentarios