Chromium VI HR
CHROMIUM VI HIGH RANGE
SPECIFICATIONS
Range 0 to 1000 µg/L
Resolution 1 µg/L
Accuracy ±5 µg/L ±4% of reading
Typical EMC ±1 µg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1687-92
,
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93723-0 Powder reagent 1 packet
REAGENT SETS
HI 93723-01 Reagents for 100 tests
HI 93723-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Chromium VI HR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette up to the mark with 10 mL of unreacted sample and replace
the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The meter will show “-0.0-” when the meter is zeroed
and ready for measurement.
Remove the cuvette and add the content of one packet of
HI 93723-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.
10 mL
Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press READ. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of total chlorine.
Note: free and total chlorine have to be measured separately with fresh unreacted samples following the
related procedure if both values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately 2
minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.