Copper HR
COPPER HIGH RANGE
SPECIFICATIONS
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.02 mg/L ±4% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the
EPA method.
The reaction between copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93702-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93702-01 Reagents for 100 tests
HI 93702-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Copper HR
method using the procedure described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The meter will show “-0.0-” when the meter is zeroed
and ready for measurement.
10 ml
Color of Water
# 3
Fill the second cuvette up to the mark with unfiltered sample and
replace the cap. This is the apparent color.
Filter 10 mL of sample through a filter with a 0.45 µm membrane into the
third cuvette, up to the 10 mL mark and replace the cap. This is the true
color.
Insert the apparent color cuvette (# 2) into the instrument and
close the lid.
Press READ to start the reading.
The meter displays the value of apparent color in PCU.
Remove the cuvette, insert the true color cuvette (# 3) into the
instrument and ensure that the notch on the cap is positioned
securely into the groove.
Press READ to start the reading. The meter displays the value of true color in PCU.
10 mL
# 2
# 3
# 2