42 43
Nickel HR
Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior
to the measurement or, alternatively, wait for 1 minute and
press READ. When the timer ends the meter will perform the
reading.
The instrument displays concentration in g/L of nickel.
INTERFERENCES
Interference may be caused by copper.
Nickel HR
SPECIFICATIONS
Range 0.00 to 7.00 g/L
Resolution 0.01 g/L
Accuracy ±4% of reading
Typical EMC ±0.02 g/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the photometric method. The reaction between nickel and the reagent
causes a blue tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93726-0 Powder reagent 1 packet
REAGENT SETS
HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Nickel HR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette up to the mark with 10 mL of unreacted
sample and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the meter
is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet of HI
93726-0 reagent. Replace the cap and shake gently until
completely dissolved.
10 mL
NICKEL HIGH RANGE