Nickel LR
Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, wait
for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix
until completely dissolved.
Fill one cuvette up to the mark with 10 mL of the blank.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the
meter is zeroed and ready for measurement.
Fill a second cuvette up to the mark with 10 mL of the reacted sample.
Insert the second cuvette into the instrument.
Press READ to start the reading. The instrument dis-
plays the results in mg/L of nickel.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings,
add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity is removed.
10 mL
10 mL
Nickel LR
NICKEL LOW RANGE
SPECIFICATIONS
Range 0.000 to 1.000 mg/L
Resolution 0.001 mg/L
Accuracy ±0.010 mg/L ±7% of reading
Typical EMC ±0.001 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 575 nm
Method Adaptation of the PAN method. The reaction between nickel and the reagents causes
an orange tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93740A-0 Phthalate-phosphate 2 packets
HI 93740B-0 0.3% PAN indicator 2 mL
HI 93740C-0 EDTA 2 packets
HI 93703-51 Dispersing Agent 4-6 drops (only when necessary, see note)
REAGENT SETS
HI 93740-01 Reagents for 50 tests
HI 93740-03 Reagents for 150 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Nickel LR
method using the procedure described in the
Method Selection
section (see page 12).
Note: for best results perform your tests between 20-24°C.
Fill one graduated beaker with 25 mL of deionized water (blank) and
another one with 25 mL of sample.
Add the content of one packet of HI 93740A-0 Phthalate-phosphate reagent
to each beaker. Cap and swirl gently until the reagent is dissolved.
Note: If sample contains iron (Fe
3+
), it is important that all powder be
dissolved completely before continuing with following step.
Add 1 mL of HI 93740B-0 0.3% PAN solution to each beaker, cap and swirl
to mix.