62 63
Silica
SILICA
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology,
D859,
Heteropoly Blue method. The reaction between silica and reagents causes a blue tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93705A-0 Molybdate 6 drops
HI 93705B-0 Citric acid 1 packet
HI 93705C-0 Amino acid 1 packet
REAGENT SETS
HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Silica
method using the procedure described
in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to
the mark).
Add 6 drops of HI 93705A-0 Molybdate reagent.
Replace the cap and swirl the solution.
Wait for 4 minutes, add the content of one packet of HI
93705B-0 Citric acid reagent and shake until it is completely
dissolved.
Wait for 1 minute. This is the blank.
10 mL
x 6
12
3
6
9
1’
Phosphorus
Add the content of one packet of HI 93706B-0 Phosphorus
Reagent B (Amino Acid) to the cuvette. Replace the cap
and shake gently until completely dissolved.
Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown
prior to the measurement or, alternatively, wait for 5
minutes and press READ. When the timer ends the
meter will perform the reading. The instrument displays
the results in mg/L of phosphorus (P).
Press the or to access the second level of functions.
Press the Chem Frm functional key to convert the result in mg/L of phosphate (PO
4
3-
) and phosphorus
pentoxide (P
2
O
5
).
Press the or to go back to the measurement screen.
INTERFERENCES
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L