Silica
SPECIFICATIONS
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading at 25 °C
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology,
D859,
Heteropoly Blue method. The reaction between silica and reagents causes a blue tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93705A-0 Molybdate 6 drops
HI 93705B-0 Citric acid 1 packet
HI 93705C-0 Amino acid 1 packet
REAGENT SETS
HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
For other accessories see page 66.
MEASUREMENT PROCEDURE
Select the
Silica
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark).
Add 6 drops of HI 93705A-0 Molybdate reagent. Replace
the cap and swirl the solution.
Wait for 4 minutes, add the content of one packet of
HI 93705B-0 Citric acid reagent and shake until it is
completely dissolved.
Wait for 1 minute. This is the blank.
Place the cuvette into the holder and close the lid.
10 mL
SILICA
Vista de pagina 55