Hanna-instruments HI 83215-2008 Series Manual de usuario Pagina 44

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 52
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 43
44
Potassium MR
POTASSIUM MEDIUM RANGE
SPECIFICATIONS
Range 10 to 100 mg/L
Resolution 2.5 mg/L
Accuracy ±15 mg/L ±7% of reading at 25 °C
Typical EMC ±2.5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 610 nm
Method Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93750A-0 Potassium Reagent 6 drops
HI 93750B-0 Powder Reagent 1 packet
REAGENT SETS
HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION (MR) procedure at page 20.
Select the
Potassium MR
method using the procedure
described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of sample, up to the mark.
Add six drops of HI 93750A-0, replace the cap and swirl
the solution.
Place the cuvette into the holder and close the lid.
10 mL