Press the Zero key. The display will show -0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one packet
of HI 93750B-0 reagent. Replace the cap and gently mix
for one minute by slowly turning the cuvette upside down.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior
to the measurement or, alternatively, wait for 2 minutes
and press Read. When the timer ends the meter will
perform the reading. The instrument displays the results in
mg/L (ppm) of potassium (K).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of potassium oxide (K
2
O).
Press or to return to the measurement screen.
INTERFERENCES
Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
Potassium MR