The
(unstable measurement) symbol will be shown on the display until the reading
becomes stable.
• When the reading is stable and close to the selected buffer, the
(unstable measurement)
symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration.
Press ESC to exit calibration.
Point 2 calibration
• The calibrated value will be shown on the display and the second expected buffer value will
be displayed.
• Remove the electrode holder with electrodes from the top of the beaker.
Place the second buffer into beaker and place in beaker holder. Rinse the electrodes in a
beaker containing the second buffer rinsing solution.
Place the electrode holder (with electrodes) on the top of the beaker, lock holder by turning.
• If necessary press the ARROW keys in order to select a different buffer value.
The
(unstable measurement) symbol will be shown on the display until the reading
becomes stable.
• When the reading is stable and close to the selected buffer, the
(unstable measurement)
symbol will disappear and the Confirm key will become active.
Press Confirm to confirm the calibration.
The calibrated value will be shown on the display and the third expected buffer value will be
automatically selected.
• After the second calibration point has been confirmed, press ESC to exit without performing
the third calibration point.