Hanna-instruments HI 84432 Manual de usuario Pagina 24

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 23
24
PURGE
Purging should be performed:
if you notice there is no titrant in the tip;
whenever the tube of the peristaltic pump is replaced;
whenever the titrant is changed or a new bottle is used;
before starting a pump calibration;
before starting a lot of titrations.
In order to start purging, press the Purge key from the titrator main screen. The purging stops
automatically after 5 minutes.
To access the Purge key follow the next steps.
• From the instrument main screen (pH meter screen) press “Titrator” function key.
The instrument will display the next screen if any of the following conditions exist:
- the meter hasn’t been calibrated in 8.20 pH buffer
- the pH calibration has expired
- a pump calibration hasn’t been performed or more than 3 days have passed since the last
pump calibration.
Press CAL to access the titration calibration menu where electrode and pump calibration may
be accessed.
Press HELP to view the contextual help.
Press Continue or ESC to skip the message and enter titrator main screen.
Vista de pagina 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 43 44

Comentarios a estos manuales

Sin comentarios