Hanna-instruments HI 84500 Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 36
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
18
Note: The aspiration tube must be inserted in the titrant bottle. The dosing tip must be placed
over a rinse beaker.
Before starting the prime procedure, make sure you are using the appropriate titrant
solution for the selected range.
To prime the burette, select Prime option from Titration mode.
Adjust the rinses number by pressing the and keys and press Start.
The number of syringe rinses can be set between 1 and 5 (at least three rinses are recommended
to ensure that the air bubbles are completely removed).
To pause the prime process press the Pause key; to continue press the Continue key. To stop the
prime process press the Stop key.
Note: This error message appears when the pump is
not working properly. Check the tubing, valve
and syringe. Press Restart to try again.
Vista de pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 35 36

Comentarios a estos manuales

Sin comentarios