For best accuracy, before taking any measurement, ensure that the pump is calibrated on the
selected range following the “Pump Calibration Procedure” (see page 19).
Select the corresponding range according to the table below:
TOTAL SO
2
MEASUREMENT PROCEDURE
egnaRwoL
)elpmasLm05(
egnaRhgiH
)elpmasLm05(
mpp0.04ot0.1 mpp004ot03
Low Range (Free & Total SO
2
) - 50 mL
High Range (Free & Total SO
2
) - 50 mL
Note: The volume of wine added is critical to the measurement
accuracy. Pipettes are recommended.
Failure to use a clean pipette will result in erroneous readings.
Fill the 20 mL beaker up to the 5 mL mark with the HI 84500-61
Alkaline Reagent, add the contents to the 100 mL beaker containing
the sample.
Cover the beaker, swirl and wait for 10 minutes.
Fill the 20 mL beaker up to the 5 mL mark with the HI 84500-60
Acid Reagent, add the contents to the 100 mL beaker containing
the sample.
Add the contents of one packet of HI 84500-62 Stabilizer Packet
to the 100 mL beaker.
Place the beaker into the beaker holder.
Press Titrator.
Ensure the pump is primed with the correct titrant for the selected range (HI 84500-50 Low
Range Titrant or HI 84500-51 High Range Titrant).
Sample preparation: Use a clean pipette to precisely add the appropriate amount of wine
sample to a clean 100 mL beaker as indicated below:
This error message appears when the pump is not working properly. Check the tubing, valve
and syringe. Press Restart to try again.