Hanna-instruments HI 96702 Manual de usuario Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
14
Remove the cuvette.
Add the content of one packet of
HI 93702-0 reagent. Replace the
cap and shake gently.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 45 seconds then
just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
copper on the LCD.
INTERFERENCE
Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted
between 6 and 8.
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 23 24

Comentarios a estos manuales

Sin comentarios