After a few seconds the display will
show “-0.0-”. The meter is now zeroed
and ready for calibration. The blinking
“READ” asks for reading calibration
standard.
Remove the cuvette.
Place the CAL CHECK™ Standard
HI 96702-11 Cuvette B into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press READ/ /TIMER and the lamp,
cuvette and detector icons will appear on
the display, depending on the
measurement phase.
After measurement the instrument will
show for three seconds the CAL CHECK™
Standard value.
Note:If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the
standard value was too low. Verify
that both CAL CHECK™ Standards
HI 96702-11 Cuvettes, A and B are
free from fingerprints or dirt and that
they are inserted correctly.
Then the date of the last calibration (e.g.:
”01.08.2009”) appears on the display,
or “01.01.2009” if the factory calibration
was selected before. In both cases the
year number is blinking, ready for date
input.