19
A 1/2 BSP sample input fitting allows direct connection to the
optional input filter.
Sample line pressure should be between 0.07 and 4 bar (1
and 57.2 psi) with an ideal pressure of 0.7 bar (10 psi).
It is recommended to assure that the water inlet come at 1 m
above the instrument sample input pitting. For maintenance
issues is also recommended to install valves on both sam-
pling pipes.
Drain Line Installation
The drain hose fitting is a 20 mm (3/4”) hose barb on the
bottom of the instrument enclosure. An air gap between the end
of the drain hose and the drain is recommended to prevent any
back flow into the instrument in the event of drain blockage.
ORP probe
PH probe
PCA 301
REAGENTS
HI 70460
TOTALCHLORINE INDICATOR SOLUTION
R:--
S:--
ETICPCA30A
Contents:
p-ToluenesulfonicAcid (6192-52-5)
Water(7732-18-5)
FORLABORATORYAND INDUSTRIAL USEONLY
DONOT FREEZE
STOREAT ROOMTEMPERATURE
KEEPCONTAINER TIGHTLYCLOSED
PCA 301
REAGENTS
HI 70461
TOTALCHLORINE BUFFER SOLUTION
Xn:Harmful
R:22
S:46
ETICPCA30B
FORLABORATORYAND INDUSTRIAL USEONLY
DONOT FREEZE
STOREATROOM TEMPERATURE
KEEPCONTAINER TIGHTLYCLOSED
HARMFULIF SWALLOWED.
IFSWALLOWED SEEKMEDICAL ADVICEIMMEDIATELY
ANDSHOW THISCONTAINER ORLABEL.
Contents:
LithiumBimaleate (unknown)
LithiumMaleate (unknown)
Water(7732-18-5)
TO DRAIN
FILTER
WATER INLET
WATER OUTLET
MIN.
1m