Hanna-instruments HI 83207-01 Manual de usuario Pagina 33

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 68
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 32
33
SPECIFICATIONS
Range 0.0 to 20.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.2 mg/L ±3% of reading at 25 °C
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition,
Periodate method. The reaction between manganese and reagents causes
a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93709A-0 Citrate 1 packet
HI 93709B-0 Sodium periodate 1 packet
REAGENT SETS
HI 93709-01 Reagents for 100 tests
HI 93709-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
Select the
Manganese HR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the cuvette.
Add the content of one packet of HI 93709A-0 Citrate reagent.
Replace the cap and invert to mix with gently movements for
2 minutes.
Manganese HR
MANGANESE HIGH RANGE
10 mL
Vista de pagina 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 67 68

Comentarios a estos manuales

Sin comentarios