Hanna-instruments HI 83207-01 Manual de usuario Pagina 38

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 68
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 37
38
Molybdenum
MOLYBDENUM
SPECIFICATIONS
Range 0.0 to 40.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±5% of reading at 25 °C
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93730A-0 Reagent A 1 packet
HI 93730B-0 Reagent B 1 packet
HI 93730C-0 Reagent C 1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
Select the
Molybdenum
method using the procedure
described in the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 67 68

Comentarios a estos manuales

Sin comentarios