Hanna-instruments HI 83207-01 Manual de usuario Pagina 34

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 68
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 33
34
Add the content of one packet of HI 93709B-0 Sodium
Periodate reagent. Replace the cap and invert to mix with
gently movements for 2 minutes.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and 30 seconds and press Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of manganese.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of potassium permanganate (KMnO
4
) and
permanganate (MnO
4
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L.
Manganese HR
Vista de pagina 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 67 68

Comentarios a estos manuales

Sin comentarios