35
pH
CLEAR
PGM
0
RANGE
5
PGM
pH
PROGRAM # 5 (HI 9017)
Setting of endpoint and alert pH alarm values.
When the alert value is reached i.e. between the final value and the
alert value, there will be a single alert beep.
When the endpoint value is reached, the
alarm will be activated continuously and
the value will simultaneously flash.
To clear the alert or alarm, press the CLEAR
key.
To stop:
• press 0
• press PGM.
E.g. activation of alarm for an endpoint value of pH 9.32 and alert
value of pH 9.03.
• set the instrument to the pH mode
• press 5
• press PGM.
At this stage, the left display shows "P5-1" and the right display
shows the measured pH value.
• press 9
• press decimal point
• press 3
• press 2
• press ENTER A.
ENTER
A
.
9
32
pH