Hanna-instruments HI 9017 Manual de usuario Pagina 8

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 64
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 7
8
SPECIFICATIONS HI9218 HI9219
Ranges pH 0.00 to 14.00
°C -10.0 to 125.0
Resolution pH 0.01
°C 0.1
Accuracy pH ±0.01
(@20°C/68°F) °C ±0.5
Typical EMC pH ±0.04
Deviation °C ±1.5
Calibration 2 point with automatic recognition of
3 buffers (pH 4.01, 7.01, 10.01)
Temperature -10° to 125°C (14 to 257°F) Completely
Compensation automatic with HI 7669/2W temperature
probe; manual by means of keyboard
Comp. Interface RS232C
Electrode HI9218: FC100B plastic body combination
pH electrode (included)
HI9219:HI1131B Glass body combination
pH electrode (included)
Temperature HI7669/2W Temperature probe (included)
Probe
Input Impedance 10
12
Ohm
Display 2 large 4-digit LCD for pH and temperature
with symbols 1 small LCD for date/time and
stability indicator
Printer 1 x 16 line dot matrix printer,
paper width: 44 mm,
roll diameter: 35 mm
Power Supply Power socket for 12VDC
Environment 0 to 50°C (32 to 122°F); RH 95%
Dimensions 230x170x70mm (9.1x6.7x2.7")
Weight 1.3 Kg (2.9 lb.);
kit w/holder 3.2 Kg (7 lb.)
DATE/TIME key to display date and time.
CAL key to enter calibration mode.
CFM key to confirm memorized buffers 7.01 and 4.01/10.01.
UP/DOWN key to set manual temperature compensation and date/time.
PAPER key manual paper feed control.
PRINT key prints pH value together with temperature, date
and time.
READY key stability indicator to ensure correct pH reading or
to enable calibration.
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 63 64

Comentarios a estos manuales