Hanna-instruments HI 903 Manual de usuario Pagina 112

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 222
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 111
7-6
AUXILIARY FUNCTIONS
The manual dispense volume must be between the limits shown below:
0.001 to 4.500 mL for a 5-mL burette
7.2.4 Purge Burette
This option allows the burette to be emptied before cleaning and/or storing.
Note: Before starting this operation, remove the aspiration tube from the titrant bottle.
The figures below show the steps in a purge burette operation.
7.3 Stirrer
Note: When custom stirrer is selected (see Section 4.4, Stirrer in General Options chapter),
the commands related to the stirrer are not available.
The stirrer can be turned on and off by pressing while in
Idle
mode.
During the titration process the stirrer cannot be turned off.
The stirring speed is set within the method parameters (
see Section,
5.5.6 Stirring Speed
).
During the titration process, the stirring speed can be manually adjusted by using the and
keys.