5-25
METHODS
5.5.15.1 Maximum Duration
Specify the maximum time a titration is allowed to run. Once this point is reached the
titration will be terminated even if the end point is not reached.
The time can be set from 10 to 3600 seconds.
5.5.15.2 Maximum Titrant Volume
The maximum titrant volume used in the titration must be set according to the analysis.
If the titration end point is not reached, the titration will be terminated after the maximum
titrant volume has been dispensed. The error message (“Limits Exceeded”) will appear on
the display.
Range is from 0.100 to 50.000 mL.