METHODS
5.5.14.1 Start Mode
The user can select the starting mode for the titration. In
Cautious
mode, the titrant dosing
begins with the minimum dose in order to prevent over-titration. In
Normal
mode, the
titrant dosing begins with the median value between the minimum and maximum (i.e.
minimum dose 5 µL, maximum dose 25 µL, first dose will be 15 µL).
5.5.14.2 Standby Mode
When enabling this option the titrator will automatically revert to Standby mode after a
titration is completed.
See also the
Standby Duration
option.