1Instruction ManualHI 83212Multiparameter BenchPhotometerfor Power Plant Utilitieswww.hannainst.comThis Instrument is inCompliance with the CE Directi

Pagina 2 - TABLE OF CONTENTS

18 19• Remove the cuvette.• Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawateranalysis). Replace the cap and mix the solution.• Add 4 drop

Pagina 3 - PRECISION AND ACCURACY

20 21FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3%

Pagina 4 - FUNCTIONAL DESCRIPTION

22 23• Reinsert the cuvette into the instrument.• Press READ to start the reading. The instrument displays the resultsin mg/L of free chlorine.INTERFE

Pagina 5 - USING CUVETTES

24 25Total Chlorine• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen

Pagina 6 - HEALTH & SAFETY

26 27Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviat

Pagina 7 - OPERATIONAL GUIDE

28 29Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 µg/LResolution 1 µg/LAccuracy ±10 µg/L ±5% of readingTypical EMC ±1 µg/LDeviationLight Sou

Pagina 8 - SETUPSETUP

30 31HydrazineHYDRAZINESPECIFICATIONSRange 0 to 400 µg/LResolution 1 µg/LAccuracy ±4% of full scaleTypical EMC ±2 µg/LDeviationLight Source Tungsten l

Pagina 9 - HELP MODE

32 33 MolybdenumMOLYBDENUMSPECIFICATIONSRange 0.0 to 40.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight

Pagina 10 - AMMONIA LOW RANGE

34 35Molybdenum• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, waitfor five minutes and press READ.

Pagina 11 - FREE CHLORINE

36 37Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until completely

Pagina 12 - INTERFERENCES

2 3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment,

Pagina 13 - TOTAL CHLORINE

38 39Phosphate LR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, alternat

Pagina 14 - COPPER HIGH RANGE

40 41Phosphorus• Add the content of one packet of HI 93706B-0 PhosphorusReagent B (Amino Acid) to the cuvette. Replace the capand shake gently until c

Pagina 15 - COPPER LOW RANGE

42 43Silica• Place the cuvette into the holder and close the lid.• Press ZERO key. The display will show “-0.0-” when themeter is zeroed and ready for

Pagina 16 - HYDRAZINE

44 45SilverSILVERSPECIFICATIONSRange 0.000 to 1.000 mg/LResolution 0.001 mg/LAccuracy ±0.005 mg/L ±10% of readingTypical EMC ± 0.001 mg/LDeviationLigh

Pagina 17 - MOLYBDENUM

46 47Silver• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Fill a second cuvette up to the mark w

Pagina 18

48 49The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software.DATA MANAGEMENTThe instrument shows clear warning me

Pagina 19 - PHOSPHATE HIGH RANGE

50 51WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose a

Pagina 20 - PHOSPHATE LOW RANGE

52Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699e-mail: [email protected]:Tel. (10) 88570068 • Fax (10) 88570060e-mail: hannachina@vi

Pagina 21 - PHOSPHORUS

4 5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wave

Pagina 22 - MEASUREMENT PROCEDURE

6 7INSTRUMENT DESCRIPTION1) Cuvette Lid2) Liquid Crystal Display (LCD).3) Splash proof keypad.4) ON/OFF power switch5) Power input connector6) USB con

Pagina 23

8 9USING LIQUID AND POWDER REAGENTS• Proper use of the dropper:(a) for reproducible results, tap the dropper on the table for several times and wipe

Pagina 24

10 11Pay attention to push the cuvette completely down in the holder.• In order to avoid reagent leaking and to obtain more accurate measure-ments, cl

Pagina 25 - STANDARD METHODS

12 13POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note:

Pagina 26 - ACCESSORIES

14 15Date / TimeThis option is used to set the instrument’s date and time.Press Modify functional key to change the date/time.Press the ⊳u functional

Pagina 27 - MAN83212 07/08

16 17Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source