32 33 Molybdenum
MOLYBDENUM
SPECIFICATIONS
Range 0.0 to 40.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±5% of reading
Typical EMC ±0.1 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93730A-0 Reagent A 1 packet
HI 93730B-0 Reagent B 1 packet
HI 93730C-0 Reagent C 1 packet
REAGENT SETS
HI 93730-01 Reagents for 100 tests
HI 93730-03 Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Select the
Molybdenum
method using the procedure
described in the
Method Selection
section (see page
12).
Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when
the meter is zeroed and ready for measurement.
Add 12 drops of the HI 93704-0 reagent to each
cuvette. Replace the caps and shake gently to mix.
Place the blank (#1) into the holder and close the lid.
Press TIMER and the display will show the countdown
prior to zeroing the blank. The display will show “-0.0-”
when the meter is zeroed and ready for measurement.
Remove the blank.
Insert the cuvette with the reacted sample (# 2) into the
instrument and close the lid.
Press READ to start the reading. The instrument displays
concentration in µg/L of hydrazine.
Hydrazine
x 12
# 1
# 2
x 12
# 1
# 2
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines