Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Fill a second cuvette up to the mark with 10 mL of the reacted sample.
Insert the second cuvette into the instrument.
Press READ to start the reading. The instrument displays
the results in mg/L of silver.
Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings,
add 2-3 drops of Dispersing agent (HI 93703-51) to each cuvette and swirl until turbidity is completely
dissolved.
INTERFERENCES
Interference may be caused by:
Al
3+
above 30 mg/L Fe
2+
above 1.5 mg/L
Ca
2+
above 1000 mg/L as CaCO
3
Fe
3+
above 10 mg/L
Cd
2+
above 20 mg/L K
+
above 500 mg/L
Cl
-
above 8000 mg/L Mn
2+
above 25 mg/L
Co
2+
above 1.5 mg/L Mg
2+
above 1000 mg/L as CaCO
3
Cr
3+
above 20 mg/L Na
+
above 5000 mg/L
Cr
6+
above 40 mg/L Ni
2+
above 1.5 mg/L
Cu
2+
above 15 mg/L Pb
2+
above 20 mg/L
F
-
above 20 mg/L Zn
2+
above 30 mg/L
10 mL
# 2
# 2
Silver
Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker
(the sample) and swirl gently to mix.
Press TIMER and the display will show the countdown prior to adding reagent C or, alternatively, wait
for 2 minutes. Then add exactly 1.0 mL of HI 93737C-0 Indicator reagent to each beaker and swirl.
Press TIMER or, alternatively, wait for 2 minutes. Then, in both cases, add 1.0 mL of HI 93737D-0
Fixing reagent to each beaker and swirl.
Then press TIMER or, alternatively, wait for 2 minutes to allow reaction to complete.
Fill one cuvette up to the mark with 10 mL of the blank.
Place the cuvette into the holder and close the lid.
# 2
# 2
# 1
# 2
10 mL
# 1
# 1
# 1
10 mL