Press RANGE/GLP/ to change the range.
The range can be changed any time when the
instrument is in
measurement mode
.
The selected range is memorized and the selection remaines unchanged at power off or when
the battery is removed.
The HI 96101 can measure pH when range P1 is selected, Free Chlorine when range P2 is
selected, Total Chlorine when range P3, Cyanuric Acid when range P4, Iodine when range P5,
Bromine when range P6, Iron LR when range P7 are selected. To change the active range follow
the procedure:
Turn the meter on by pressing ON/OFF. The
display briefly shows all tags on.
After startup, the range identification number is
displayed on the secondary LCD as P1, P2, P3,
P4, P5, P6 or P7.
RANGE SELECTION
Code Parameter
P1 pH
P2 Free Chlorine
P3 Total Chlorine
P4 Cyanuric Acid
P5 Iodine
P6 Bromine
P7 Iron LR
MEASUREMENT PROCEDURE
To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.
Important note: Free and Total Chlorine have to be measured separately with fresh unreacted
samples if both values are requested.
Turn the meter on by pressing ON/OFF.
The display briefly shows all tags on.