Fill a beaker up to the 25 mL mark with
the sample, add the content of one packet
of HI 93722-0 reagent and swirl gently
to mix. This is the reacted sample.
Fill a second cuvette with 10 mL of the
reacted sample, up to the mark, and
replace the cap.
Place the second cuvette into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
Alternatively, wait for 45 seconds then
just press READ/ /TIMER. In both
cases, the lamp, cuvette and detector
icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the
instrument directly displays concentration
in mg/L of cyanuric acid on the LCD.