Hanna-instruments HI 96101 Manual de usuario Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 36
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
23
IODINE MEASUREMENT
Fill the cuvette with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
After a few seconds, the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
Remove the cuvette.
Add the content of one packet of
HI 93718-0 Iodine reagent. Replace the
cap and shake gently for about 20 seconds
to dissolve most of the reagent.
Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.
Vista de pagina 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 35 36

Comentarios a estos manuales

Sin comentarios