Hanna-instruments HI 96101 Manual de usuario Pagina 26

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 36
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 25
26
IRON LR MEASUREMENT
Fill one graduating mixing cylinder up to the 25
mL mark with deionized water.
Add the content of one packet of HI 93746-0
TPTZ reagent, close the cylinder and shake
vigorously for 30 seconds. This is the blank.
Fill the cuvette with 10 mL of the blank up to the
mark and replace the cap.
Place the cuvette into the cuvette holder and ensure
that the notch on the cap is positioned securely
into the groove.
Press ZERO/CFM and the lamp, cuvette and
detector icons will appear on the display, depending
on the measurement phase.
After a few seconds, the display will show “-0.0-”.
The meter is now zeroed and ready for measurement.
Remove the cuvette.
Fill another graduated mixing cylinder up to the
25 mL mark with the sample.
Add the content of one packet of HI 93746-0
TPTZ reagent, close the cylinder and shake vigor-
ously for 30 seconds. This is the reacted sample.
Fill a cuvette with 10 mL of the reacted sample up
to the mark and replace the cap.
25mL25mL
25mL25mL
25mL
12
3
6
9
30"
25mL25mL
25mL25mL
25mL
12
3
6
9
30"
Vista de pagina 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 35 36

Comentarios a estos manuales

Sin comentarios