26
SINGLE SAMPLE READ
• Remove the cuvette from the instrument.
• Remove the cap.
• Add the content of one packet of the specific
test reagent, for:
Free Chlorine Total Chlorine
1 packet of or 1 packet of
HI 93701-0
HI 93711-0
• Replace the cap and shake gently for 20 seconds
(or 2minutes in case of seawater analysis).
• Replace the cuvette into the holder and ensure
that the mark on the glass is matched with the
mark on the instrument top. Close the lid.
• Hold READ/TIMER u for 3 seconds. The
display will show the hourglass blinking and
the countdown prior to measurement.
Alternatively wait for:
Free Chlorine Total Chlorine
1 minute or 2 minutes and 30 seconds
and then just press READ/TIMER u.
• In both cases blinking dashes will appear during
measurement.
• The instrument directly displays the concentration
in mg/L of free or total chlorine.
Note:If the value is over range, the maximum
value (5.00 mg/L) will blink.