MULTIPLE SAMPLES READ
Place the second cuvette with the reacted
sample into the holder and ensure that the
mark on the glass is matched with the mark on
the instrument top.
Hold READ/TIMER u for 3 seconds. The
display will show the hourglass blinking and
the countdown prior to measurement.
Alternatively wait for:
Free Chlorine Total Chlorine
1 minute or 2 minutes and 30 seconds
and then just press READ/TIMER u.
In both cases blinking dashes will appear during
measurement.
The instrument directly displays the concentration
in mg/L of free or total chlorine.
Note:It is recommended to make a zero before
each measurement.
Note:If the signal to noise ratio is too high, the
zero value will blink.
INTERFERENCES
The colorimetric measurements are affected by the following interfering agents:
Bromine (positive error).
Chlorine dioxide (positive error).
Iodine (positive error).
Oxidized Manganese and Chromium (positive error).
Alkalinity above 250 mg/L CaCO
3
or acidity above 150 mg/L CaCO
3
will not reliably develop the
full amount of color or it may rapidly fade. To resolve this, neutralize the sample with diluted HCl
or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximatively
2 minutes after adding the powder reagent.