Turn the instrument on by pressing ON/OFF.
Make sure that the instrument is in the free or
total chlorine range (the desired one).
Press RANGE s to select the desired range
(free or total chlorine). When dashes appear
on the LCD, the instrument is ready. The
ZERO” tag will blink on the LCD.
Place the zero cuvette (A) in the instrument
with the mark aligned with the mark on the
instrument top.
Close the lid and press ZERO/AVG t. The
LCD will display blinking dashes and the
measuring icon during zero measurement. At
the end of zero measurement the “-0.0-” is
displayed. The meter is now ready for
validation.
COLORIMETER CALIBRATION
The HI 93414 free and total chlorine colorimeter has a powerful CAL CHECK function that allows the
user to check the instrument calibration against a NIST traceable standard before making a set of
measurements. With the same standard, the instrument could be re-calibrated, if necessary.
Note:Free and total chlorine must be calibrated separately. Calibration of one range will not
calibrate the other range.
VALIDATION PROCEDURE
Warning: Do not validate or calibrate the instrument with standard solutions other than Hanna CAL
CHECK™ Standards, otherwise erroneous results will be obtained. For accurate validation
and calibration please perform test at room temperature, 18 to 25 ºC (64.5 to 77.0 ºF).