Close the lid and press READ/TIMER u. The
display will show blinking dashes and the icons
for cuvette, detectors and lamp will appear
during measurement.
At the end of the measurement, the third
calibration point (100 NTU) is displayed on
the primary LCD and “CAL P.3” on the
secondary LCD and “READ” tag will blink.
At this moment it is possible to exit calibration
by pressing CAL/CHECK. The instrument will
memorize the two-point calibration data and
will return to measurement mode.
THREE-POINT CALIBRATION
To perform a three-point calibration, continue the
procedure with the following steps:
Remove the second standard cuvette.
Place the 100 NTU standard cuvette (or the
third prepared formazin standard) into the
holder, with the cuvette mark aligned with the
sign on the instrument top.
Close the lid and press READ/TIMER u. The
display will show blinking dashes and the icons
for cuvette, detectors and lamp will appear
during measurement.
At the end of the measurement, the fourth
calibration point (750 NTU) is displayed on
the primary LCD and “CAL P.4” on the
secondary LCD and “READ” tag will blink.
At this moment it is possible to exit calibration
by pressing CAL/CHECK. The instrument will
memorize the three-point calibration data and
will return to measurement mode.