FOUR-POINT CALIBRATION
To perform a four-point calibration, continue the
procedure with the following steps:
Remove the third standard cuvette.
Place the 750 NTU standard cuvette (or the
fourth prepared formazin standard) into the
holder, with the cuvette mark aligned with the
sign on the instrument top.
Close the lid and press READ/TIMER u. The
display will show blinking dashes and the icons
for cuvette, detectors and lamp will appear
during measurement.
At the end of the measurement, the four-point
calibration is completed and the instrument
returns automatically to measurement mode.
OUT CAL RANGE FUNCTION
The instrument has an Out Cal Range function to
prevent taking measurements in a range where the
calibration do not assure the best results. The range
where the calibration assures correct measurements
is up to 40 NTU for a two-point calibration and up
to 150% of the third point value for a three-point
calibration.
The display will show a blinking “CAL” tag each
time the measurements are taken outside the
calibration range.
CALIBRATION ERROR MESSAGES
If the value of the standard read during the
calibration is too far from the set value, the
instrument will display “-LO-” or “-HI-” error
messages. Check if the correct standard is used
or prepare a fresh standard, if formazine is
used, and repeat the reading of the standard.