Instruction Manual
HI 83214
Multiparameter Bench
Photometer
for Wastewater Treatment
www.hannainst.com
Vista de pagina 0
1 2 3 4 5 6 ... 71 72

Indice de contenidos

Pagina 1 - HI 83214

1Instruction ManualHI 83214Multiparameter BenchPhotometerfor Wastewater Treatmentwww.hannainst.com

Pagina 2 - TABLE OF CONTENTS

10(b)(c)(a)invert to mix• Proper use of the powder reagent packet:(a) use scissors to open the powder packet(b) push the edges of the packet to form a

Pagina 3 - GENERAL DESCRIPTION

11shake gently shake vigorously(b) shake the vial: move the vial up and down. The movement may be gentle or vigorous.This mixing technique is indicate

Pagina 4 - SPECIFICATIONS

12DIGESTION• Some analytical methods require digestion of the sample. Use the HI 839800 test tube heater fordigestion. The optional HI 740217 safety s

Pagina 5

13METHOD REFERENCE TABLEMethod Method Method Pagereagent set code description1 HI 93764A-25 Ammonia LR 202 HI 93764B-25 Ammonia HR 223 HI 93701-01 Chl

Pagina 6 - FUNCTIONAL DESCRIPTION

14VIALS IDENTIFICATIONVials related to different parameters can be distinguished by means of the letter printed on the vial (beside theHanna logo) and

Pagina 7 - KEYPAD DESCRIPTION

15OPERATIONAL GUIDEPOWER CONNECTIONConnect the AC/DC adaptor that is supplied with the instrument.Note: Always turn the meter off before unplugging it

Pagina 8

16DATA MANAGEMENTYou can keep track of your results using the instrument’s data log function. Up to 200 individualmeasurements can be stored in the da

Pagina 9 - Probable level of

17In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are general parameters that ch

Pagina 10 - USING VIALS

18Date / TimeThis option is used to set the instrument’s date and time.Press the Modify key to change the date/time.Press the ⊳ u functional keys to h

Pagina 11

19HI 83214 offers an interactive contextual help mode that assists the user at any moment.To access the help screens press HELP.The instrument will di

Pagina 12 - HEALTH & SAFETY

2Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using themeter. This manual will provide y

Pagina 13 - METHOD REFERENCE TABLE

20SPECIFICATIONSRange 0.00 to 3.00 mg/L (as NH3-N)Resolution 0.01 mg/LAccuracy ±0.10 mg/L or ±5 % of reading @ 25 °C, whichever is greaterLight Source

Pagina 14 - VIALS IDENTIFICATION

21• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready formeasurement.• Remove the vial.• Remove the cap and add 4 d

Pagina 15 - OPERATIONAL GUIDE

22Ammonia HRAMMONIA HIGH RANGESPECIFICATIONSRange 0 to 100 mg/L (as NH3-N)Resolution 1 mg/LAccuracy ±1 mg/L or ±5 % of reading @ 25 °C, whichever is g

Pagina 16 - Viewing and deleting

234Ammonia HR• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready formeasurement.• Remove the vial.• Remove the cap

Pagina 17 - Backlight

24CHLORINE, FREEChlorine, Free10 mLSPECIFICATIONSRange 0.00 to 5.00 mg/L Cl2Resolution 0.01 mg/LAccuracy ±0.03 mg/L or ±4 % of reading @ 25 °C, whiche

Pagina 18 - Tutorial

25• Remove the cap and add the content of one packet of HI 93701-0 FreeChlorine Reagent.• Replace the cap and shake gently to mix for about 20 seconds

Pagina 19 - HELP MODE

26CHLORINE, TOTALChlorine, Total10 mLSPECIFICATIONSRange 0.00 to 5.00 mg/L Cl2Resolution 0.01 mg/LAccuracy ±0.03 mg/L or ±4 % of reading @ 25 °C, whic

Pagina 20 - Ammonia LR

27• Remove the cap and add the content of one packet of HI 93711-0 TotalChlorine Reagent.• Replace the cap and shake gently to mix for about 20 second

Pagina 21 - INTERFERENCES

28SPECIFICATIONSRange 0.0 to 30.0 mg/L NO3--NResolution 0.1 mg/LAccuracy ±1.0 mg/L or ±5 % of reading @ 25 °C, whichever is greaterLight Source Tungst

Pagina 22 - Ammonia HR

29• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready formeasurement.• Remove the vial.• Remove the cap and add the

Pagina 23

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl

Pagina 24 - edition

30Nitrogen, Total LRNITROGEN, TOTAL LOW RANGESPECIFICATIONSRange 0.0 to 25.0 mg/L NResolution 0.1 mg/LAccuracy ±1.0 mg/L or ±5 % of reading @ 25 °C, w

Pagina 25

31BlankD.I. WaterSampleSampleBlankSample← green label →Nitrogen, Total LRMEASUREMENT PROCEDUREBefore using the reagent kit carefully read all the in

Pagina 26 - Chlorine, Total

32BlankSampleBlankSampleBlankSampleNitrogen, Total LR• Replace the cap tightly and shake the vials vigorously for about 30 seconds until all the powde

Pagina 27

33BlankDigestedblankSampleDigestedsampleBlankSampleBlankSampleNitrogen, Total LR← red label →SampleBlank• Press Start to activate the 2 minutes time

Pagina 28 - Nitrate

34• Press Start and the display will show the countdown prior to the measurement and the “ReactionTime” message. Alternatively, wait for 5 minutes an

Pagina 29

35NITROGEN, TOTAL HIGH RANGENitrogen, Total HRSPECIFICATIONSRange 10 to 150 mg/L NResolution 1 mg/LAccuracy ±3 mg/L or ±4 % of reading @ 25 °C, whiche

Pagina 30 - Reagent Vial identification

36BlankD.I. WaterSampleSampleBlankSampleMEASUREMENT PROCEDUREBefore using the reagent kit carefully read all the instructions and the Material Safety

Pagina 31 - ← green label →

37BlankSampleBlankBlankSampleNitrogen, Total HRSample• Replace the cap tightly and shake vigorously the vials for about 30 seconds until all the powde

Pagina 32 - Nitrogen, Total LR

38BlankSampleBlankDigestedblankSampleDigestedsample← green label →SampleNitrogen, Total HR• Remove the cap from the vials and add the content of one p

Pagina 33 - ← red label →

39• Press Start and the display will show the countdown prior to the measurement and the ”ReactionTime” message. Alternatively, wait for 5 minutes and

Pagina 34

4PRINCIPLE OF OPERATIONAbsorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.When a light beam cross

Pagina 35

40OXYGEN DEMAND, CHEMICAL LOW RANGESPECIFICATIONSRange 0 to 150 mg/L CODResolution 1 mg/LAccuracy ±5 mg/L or ±5 % of reading @ 25 °C, whichever is gre

Pagina 36 - D.I. Water

41BlankSampleBlankD.I. WaterSampleSample• Choose a homogeneous sample. Samples containing settleable solids need to be homogenized with ablender.• Pre

Pagina 37 - Nitrogen, Total HR

42• Select Oxygen Demand, (COD) LR method following one of the procedures described in the “MethodSelection” section (see page 15).• Place the bla

Pagina 38 - ← green label →

43OXYGEN DEMAND, CHEMICAL MEDIUM RANGESPECIFICATIONSRange 0 to 1500 mg/L CODResolution 1 mg/LAccuracy ±15 mg/L or ±4 % of reading @ 25 °C, whichever i

Pagina 39

44• Choose a homogeneous sample. Samples containing settleable solids need to be homogenized with ablender.• Preheat the Hanna Reactor HI 839800 to 15

Pagina 40

45COD MR• Select the Oxygen Demand, (COD) MR method following one of the procedures described in the“Method Selection” section (see page 15).• Place t

Pagina 41 - ← red label →

46OXYGEN DEMAND, CHEMICAL HIGH RANGESPECIFICATIONSRange 0 to 15000 mg/L CODResolution 10 mg/LAccuracy ±150 mg/L or ±3 % of reading @ 25 °C, whi

Pagina 42 - Oxygen Demand, (COD) LR

47• Choose a homogeneous sample. Samples containing settleable solids need to be homogenized with ablender.• Preheat the Hanna Reactor HI 839800 to 15

Pagina 43

48COD HR• Select the Oxygen Demand, (COD) HR method following one of the procedures described in the“Method Selection” section (see page 15).• Place t

Pagina 44 - ← white label →

49SPECIFICATIONSRange 0.00 to 1.60 mg/L PResolution 0.01 mg/LAccuracy ±0.05 mg/L or ±5 % of reading @ 25 °C, whichever is greaterLight Source

Pagina 45 - Oxygen Demand, (COD) MR

5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wavele

Pagina 46

50• Remove the cap and add the content of one packet of HI 93758-0Phosphorus Reagent.• Replace the cap tightly and shake gently to mix for 2 minutes u

Pagina 47

51PHOSPHORUS, ACID HYDROLYZABLEPhosphorus, Acid HydrolyzableSPECIFICATIONSRange 0.00 to 1.60 mg/L PResolution 0.01 mg/LAccuracy ±0.05 mg/L or

Pagina 48 - Oxygen Demand, (COD) HR

52Phosphorus, Acid Hydrolyzable• Remove the cap from a Reagent Vial.• Add exactly 5.0 mL of sample to the vial, while keeping the vial at a45-degree a

Pagina 49 - PHOSPHORUS, REACTIVE

53Phosphorus, Acid Hydrolyzable• Press the Zero key. The instrument will perform a zero sequence and after a few seconds the displaywill show “-0.0-”.

Pagina 50

54INTERFERENCESInterference may be caused by:Arsenate at any levelSilica above 50 mg/LSulfide above 9 mg/L.To eliminate sulfide: add Bromine Water dro

Pagina 51

55PHOSPHORUS, TOTAL LOW RANGESPECIFICATIONSRange 0.00 to 1.15 mg/L PResolution 0.01 mg/LAccuracy ±0.05 mg/L or ±6 % of reading @ 25 °C, whiche

Pagina 52 - Acid Hydrolyzable

56Phosphorus, Total• Preheat the Hanna Reactor HI 839800 to 150 °C (302 °F). For correct use of the reactor follow ReactorInstruction Manual.Use of th

Pagina 53 - Phosphorus, Acid Hydrolyzable

57Phosphorus, Total• Press the Zero key. The instrument will perform a zero sequence and after a few seconds the displaywill show “-0.0-”. Now the met

Pagina 54

58Note: For accurate measurements1) wash glassware only with phosphate-free detergents2) clean all glassware with 1 : 1 hydrochloric acid solution and

Pagina 55

59PHOSPHORUS, REACTIVE HIGH RANGEPhosphorus, Reactive HRSPECIFICATIONSRange 0.0 to 32.6 mg/L PResolution 0.1 mg/LAccuracy ±0.5 mg/L or ±5 % of

Pagina 56

6INSTRUMENT DESCRIPTION1) Vial holder2) Liquid Crystal Display (LCD)3) Keypad4) Bottom lid5) USB connector6) ON/OFF power switch7) Power supply connec

Pagina 57 - Phosphorus, Total

60BlankSampleBlankD.I. WaterSampleSamplePhosphorus, Reactive HR• Replace the cap and mix by inverting each vial a couple of times.• Place the blank vi

Pagina 58

61Note: For accurate measurements1) wash glassware only with phosphate-free detergents2) clean all glassware with 1 : 1 hydrochloric acid solution and

Pagina 59

62PHOSPHORUS, TOTAL HIGH RANGEPhosphorus, Total HRSPECIFICATIONSRange 0.0 to 32.6 mg/L (P)Resolution 0.1 mg/LAccuracy ±0.5 mg/L or ±5 % of rea

Pagina 60

63BlankSampleBlankSampleBlankD.I. WaterSampleSamplePhosphorus, Total HRReagent Blank Correction: This method requires a reagent blank correction. A si

Pagina 61

64BlankSampleBlankSampleBlankSampleBlankSamplePhosphorus, Total HR• Select the Phosphorus, TTTTTotal High Range method following one of the procedures

Pagina 62

65• Remove the blank vial.• Place the sample vial into the holder and push it completely down.• Press Read. The instrument will perform the reading. T

Pagina 63

66The instrument shows clear messages when erroneous conditions appear and when measured values are outsidethe expected range. These messages are desc

Pagina 64 - Total High Range

67Description Range MethodAmmonia LR 0.00 to 3.00 mg/L NesslerAmmonia HR 0 to 100 mg/L NesslerChlorine, Free 0.00 to 5.00 mg/L DPDChlorine, Total 0.00

Pagina 65

68ACCESSORIESREAGENT SETSHI 93701-01 100 free chlorine testsHI 93701-03 300 free chlorine testsHI 93711-01 100 total chlorine testsHI 93711-03 300 tot

Pagina 66 - ERRORS AND WARNINGS

69WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose and

Pagina 67 - DATA MANAGEMENT

7KEYPAD DESCRIPTIONThe keypad contains 8 direct keys and 3 functional keys with the following functions:Press to perform the function displayed above

Pagina 68 - ACCESSORIES

70Hanna Instruments reserves the right to modify the design, construction or appearance of its productswithout advance notice.Recommendations for User

Pagina 70

72Printed in ROMANIAHanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTel. (800)

Pagina 71

8TIPS FOR AN ACCURATE MEASUREMENTThe instructions listed below should be carefully followed during testing to ensure most accurate results.COLLECTING

Pagina 72 - MAN83214 05/15

9(c)(a) (b)Probable level ofliquid taken upby syringe(a) (b)2. In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:(a) push the plunge