PHOSPHORUS, REACTIVE HIGH RANGE
Phosphorus, Reactive HR
SPECIFICATIONS
Range 0.0 to 32.6 mg/L P
Resolution 0.1 mg/L
Accuracy ±0.5 mg/L or ±5 % of reading @ 25 °C, whichever is greater
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
20
th
edition
, 4500-P C, vanadomolybdophosphoric acid method. The reaction between
orthophosphate and the reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code Description Q.ty/test Q.ty/set
HI 93763A-0* Reagent Vial 1 vial 50 vials
DEIONIZED120 Deionized Water 5 mL 1 bottle
*
Reagent Vial identification
: P RHR, green label.
Note: Store the unused vials in their container in a cool and dark place.
REAGENT SET
HI 93763A-50 Reagents for up to 49 tests
For other accessories see page 68.
MEASUREMENT PROCEDURE
Reagent Blank Correction: This method requires a reagent blank correction. A single blank vial may be used
more than once; the blank vial is stable up to two weeks (room temperature). For most accurate measurement,
run a blank for each set of measurements and always use the same lot of reagents for blank and samples.
Select the
Phosphorus, Reactive High Range
method following one of the procedures described in the
“Method Selection” section (see page 15).
Remove the caps from two Reagent Vials.
Add exactly 5.0 mL of sample to one vial (sample vial), and 5.0 mL of deionized water to the other
vial (blank vial), while keeping the vials at a 45-degree angle.
green label