Press Start and the display will show the countdown prior to the measurement and the “Reaction
Time” message. Alternatively, wait for 5 minutes and press Zero.
The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the blank vial.
Place the sample vial into the holder and push it completely down.
Press Read. The instrument will perform the reading. The instrument displays concentration in mg/L of
total nitrogen (N).
Press s or t to convert the result in mg/L of ammonia (NH
3
) or mg/L of nitrate (NO
3
-
).
The method detects all organic and inorganic forms of nitrogen present in the sample.
INTERFERENCES
Interference may be caused by:
Bromide (Br
-
) above 60 mg/L
Chloride (Cl
-
) above 1000 mg/L
Chromium (Cr
3+
) above 0.5 mg/L
Nitrogen, Total LR