Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
Remove the vial.
Remove the cap and add the content of one packet of HI 93766-0 Nitrate
Reagent.
Replace the cap tightly and invert the vial 10 times. This is the reacted
sample.
Note: The method is technique sensitive: to obtain most reproducible results
it is strongly recommended to follow carefully the “invert” procedure
described on page 10.
Place the vial into the holder and push it completely down.
Press Timer and the display will show the countdown prior to the measurement and the ”Reaction
Time message. Alternatively, wait for 5 minutes and press Read. The instrument will perform the
reading.
The instrument displays concentration in mg/L of nitrate-nitrogen (NO
3
-
-N). Press s or t to
convert the result in mg/L of nitrate (NO
3
-
).
INTERFERENCES
Interference may be caused by:
Barium (Ba
2+
) above 1 mg/L
Chloride (Cl
-
) above 1000 mg/L
Nitrite (NO
2
-
) above 50 mg/L
Samples containing up to 100 mg/L nitrite may be measured after the following treatment: add 400 mg
of urea to 10 mL of sample, mix until completely dissolved, then proceed with the usual measurement
procedure.
Nitrate